The parameters for particle therapy systems from Chinese institutes and overseas vendors.

Accelerator Synchrotron Synchrotron Synchrotron Synchrotron Synchro-cyclotron Synchrocy-clotron Cyclotron Cyclotron
Institute/Vendor Shanghai Ruijin Hospital Lanzhou, IMP Shanghai Proton and Heavy IonCenter (Siemens) Hitachi IBA S2C2 Proteus ONE MeVion S250i Hyperscan IBA Proteus 235 Varian ProBeam
Particle type Proton Carbon ions Proton and carbon ions Proton and carbon ions Proton Proton Proton Proton
Energy range 70–235 MeV 70–400 MeV/u P: 48–221 MeV, C: 85–430 MeV/u P: 70–250 MeV, C: 120–480 MeV/u Max. 230/235 MeV Max. 227 MeV 70–230 MeV 70–250 MeV
Dose rate >2 Gy/min/L 1.0 Gy/min/L 2 Gy/min/L, 2nA 2 Gy/min/L
Extraction intensity 4–8 × 1010 p/spill 6 × 108 ions/spill 2 × 1010 p/spill, 5 × 108 ions/spill 2.8 × 107 p/s
Accelerator configuration FODO, 8 bends 6-fold symmetry 6 Dipoles, 10 Quadrupoles (conventional); 4 Dipoles and 4 Quadrupoles (new) Superconduting synchrocy-clotron Superconduting synchrocy-clotron Isochronous cyclotron Isochronous superconducting cyclotron
Cicumference or Diameter 24.6 m 56.2 m 64.8 m 23 m (conventional) and 18 m (new) 2.5 m 1.8 m 4.34 m 3.2 m
Injection Multi-trun Multi-trun injection by an LINAC Multi-trun injection by an LINAC
Ion source ECR 7 MeV 6.8 MeV/u, 10 uA by a cyclotron ECR 7 MeV/u ECR 7 MeV/u PIG PIG PIG PIG
RF system 1–7 MHz, Max. Voltage: 2.5 kV 1.6–8 MHz (conventional), 1.3–10 MHz (new) 90–60 MHz, Voltage: 7–10 kV 106 MHz 72.8 MHz
Repetition frequency 0.5–0.1 Hz (variable) 1 kHz 750 Hz CW CW
Extraction RF-knockout slow extraction RF-knockout slow extraction 3rd order resonance extraction RF-driven slow extraction Passive regenerative Electrostatic deflector Electrostatic deflector
Pulse duration 1–10 s 0.5–5 s 10 μs 10 μs
Energy modulation SEE and MEE SEE SEE and MEE ESS in gantry No ESS, range shifting plates in gantry ESS in BTS ESS in BTS
Magnetic Field central/extraction 1.7 T 5.75/5 T 0.9/2.188 T 2.4/4 T
Beam ON/OFF time <150 us <15 us <300 us
Energy switch time 8 s for SEE, 0.2 s for MEE 2 s for SEE, 0.2 s for MEE 50 ms
Reference [18] [19, 20] Huang and Li [7, 15, 21] [2224] [25, 26] [27, 28] [17, 29]

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.