Table 2
Comparison of the most important parameters of the accelerators currently in the use of particle therapy.
Isochronous cyclotron | Synchrocyclotron | Synchrotron | |
---|---|---|---|
Protons | Y | Y | Y |
Carbon ions | N | N | Y |
Time Structure | Continuous 2 | Pulsed, duty cycle 1% | Spill, dead time > 10% |
Fast energy modulation | Degrader | Degrader | Wait for next spill |
Activation | Shielding | Shielding | N |
Beam intensity | Any, adjustable within ms | Any, adjustable within ms | Limited in magnitude and range |
Intensity stability | 2–5% | 2–5% | 10–20% |
Spot scanning | Y | Y | Y |
Fast continuous scanning | Y | Y | Difficult due to pulse structure |