Comparison of the most important parameters of the accelerators currently in the use of particle therapy.

Isochronous cyclotron Synchrocyclotron Synchrotron
Protons Y Y Y
Carbon ions N N Y
Time Structure Continuous 2 Pulsed, duty cycle 1% Spill, dead time > 10%
Fast energy modulation Degrader Degrader Wait for next spill
Activation Shielding Shielding N
Beam intensity Any, adjustable within ms Any, adjustable within ms Limited in magnitude and range
Intensity stability 2–5% 2–5% 10–20%
Spot scanning Y Y Y
Fast continuous scanning Y Y Difficult due to pulse structure